Faulty heart defibrillator alert

 64258611 symbol on defibrillator paddle spl 1 Defibrillators deliver a shock in order to get the heart beating normally again

More than 1,300 emergency heart devices in public places like shopping centres may not work because of a battery fault, warns a UK health regulator.

The defibrillator machines are used to give life-saving treatment to patients having a cardiac arrest.

The Medicines and Healthcare products Regulatory Agency says first aiders should check their samaritan PAD 300/300P defibrillators.

They can turn on or off when not in use, draining the battery.

The following serial number ranges are affected with one or both of the identified faults:

  • 0400000501 to 0700032917 inclusive
  • 08A00035000 to 10A00070753 inclusive
  • 10C00200000 to 10C00210318 inclusive

These were distributed around the UK between August 2004 and December 2010 with a warranted life of up to seven years.

No other HeartSine Technologies Ltd automated external defibrillator products are affected.

John Wilkinson, the MHRA’s director of medical devices, said: “People who are responsible for these public access defibrillators that are in use at shopping centres, railway stations, dental surgeries and other public places, need to check the serial numbers and, if they have an affected device, follow the advice in the manufacturer’s field safety notice.”